Simultaneous determination of vitamin B6 and magnesium using capillary electrophoresis coupled with contactless conductivity detection: Method development, validation, and application to pharmaceutical and nutraceutical samples

Thi Huyen My Dang, Huy Nguyen Quang, Bach Pham, Thi Minh Thu Nguyen, Thi Minh Hien Lu, Cong Khanh Cao, Van Anh Kieu, Yen Nhi Do, Thanh Duc Mai, Dinh Chi Le, Anh Quoc Hoang, Thi Anh Huong Nguyen, Sep-2022, In: Journal of Chromatography B, 1210, p. 123471

Overview

  • Thi Huyen My Dang
  • Huy Nguyen Quang
  • Bach Pham
  • Thi Minh Thu Nguyen
  • Thi Minh Hien Lu
  • Cong Khanh Cao
  • Van Anh Kieu
  • Yen Nhi Do
  • Thanh Duc Mai
  • Dinh Chi Le
  • Anh Quoc Hoang
  • Thi Anh Huong Nguyen

Abstract:

Vitamins and minerals are usually incorporated in pharmaceutical and nutraceutical products, but a simple, rapid, and inexpensive analytical method for their simultaneous determination is still lacking. In this study, we developed a quantification method for pyridoxine (vitamin B6) and magnesium (Mg) by using purpose-made

capillary electrophoresis with capacitively coupled contactless conductivity detection (CE-C
4D) instrument. Main analytical conditions include: fused silica capillary (total length 55 cm, effective length 40 cm, inner

diameter 50
μm); background electrolyte consisted of 10 mM L-arginine/acetic acid (pH 5) with 20% acetonitrile; separation voltage + 20 kV; hydrodynamic injection (siphoning at 20 cm in 25 s). Detection limits of vitamin B6 and Mg were 1 and 0.1 mg/L, respectively. Good linearity (R2 > 0.999) was observed for vitamin B6

and Mg calibration curves over concentration ranges of 3100 and 0.3200 mg/L, respectively. The method was

applied to analyze vitamin B
6 and Mg in several pharmaceutical and nutraceutical samples. The analytical results obtained by our method were in good agreement with reference methods (i.e., HPLC for vitamin B6 and ICP-OES for Mg). High-efficient and low-cost CE-C4D method can accordingly serve as a promising tool for concurrent analysis of inorganic and organic species in pharmaceutical and nutraceutical analysis.

Keyword(s): CE-C4D Magnesium Vitamin B6 Pharmaceutical Nutraceutical

Article number 123471
Journal Journal of Chromatography B
Volume 1210
Publication status Published - Sep-2022
ISBN 1570-0232